Harmonics and Flicker Testing
Harmonics and flicker testing to IEC 61000-3-2 and IEC 61000-3-3, for CE marking and the other markets that use the same standards.
- Both Standards, One System: Harmonic current and voltage flicker are measured on one calibrated source and analyzer system, so both results come from the same booking for your technical file.
- Your Class Agreed Up Front: Class A, B, C or D sets the limits you have to meet, so we agree your class from the product ratings when we scope the test.
- On-Site Testing Available: Large or installed equipment that cannot ship can be measured at your facility.
- The Full CE EMC Package: Emissions, immunity and product safety testing run in the same lab, so harmonics and flicker are not a separate project.
Accredited EMC testing

ISO/IEC 17025
A2LA #5179.02

FCC recognized
Part 15/18/22 · wireless

ISED recognized
CAB CA0102
* Tell us your requirements for a detailed quote and project timeline. We’ll respond within 24 hours.
What are the CE requirements for harmonics and flicker testing?
For CE marking under the EMC Directive 2014/30/EU, mains-connected equipment rated up to 16 A per phase normally shows conformity by meeting the harmonized standards EN 61000-3-2 for harmonic current emissions and EN 61000-3-3 for voltage fluctuations and flicker. These standards are usually called up through your product or generic EMC standard, and the test report goes into the technical file behind your EU Declaration of Conformity.
The UK accepts the same EN standards for UKCA and CE marking. Australia and New Zealand have adopted both standards as AS/NZS editions, and South Africa as SANS editions. All are built on the IEC originals.
Which products need harmonics and flicker testing?
Anything that plugs into the public 230 V class low-voltage supply and draws up to 16 A per phase is in scope. That covers most household appliances, IT equipment, lighting, power tools, chargers, heaters, pumps and motor-driven products.
Two kinds of product need the closest attention: switch-mode supplies without power factor correction, which fail on harmonics, and heaters, compressors and motors that switch heavy loads, which fail on flicker.
Your product or generic standard decides whether the tests apply. Send us the ratings and we will confirm it.
IEC 61000-3-2 harmonic current emissions
A product with a non-linear input stage does not draw a smooth sine wave of current. The distorted current contains harmonics, multiples of the 50 or 60 Hz supply frequency, and when thousands of products on the same network do the same, those harmonics distort the voltage for everyone. IEC 61000-3-2 measures each harmonic from the 2nd to the 40th and compares it with the limit for the product’s class.
| Class | Equipment | Examples |
|---|---|---|
| Class A | Balanced three-phase equipment, household appliances and anything not in another class | Washing machines, vacuum cleaners, audio equipment |
| Class B | Portable tools and non-professional arc welding equipment | Hand drills, portable saws, hobby welders |
| Class C | Lighting equipment | LED luminaires and lamps |
| Class D | Personal computers and PC monitors, television receivers, and refrigerators and freezers with variable-speed compressor drives, at a specified power up to 600 W | Desktop PCs, monitors, televisions |
What decides your limit
The class comes first. Class A and Class B limits are fixed currents for each harmonic, while Class C limits for lighting are a percentage of the input current and Class D limits scale with the product’s power. Some low-power equipment is exempt from the limits altogether. Send us the product’s ratings and we will confirm the class and whether an exemption applies when we quote.
IEC 61000-3-3 voltage fluctuations and flicker
When a product’s load changes, the current it draws changes, and the supply voltage dips across the network impedance. Repeat that often enough and nearby lamps visibly flicker. IEC 61000-3-3 limits how much voltage change a product may cause, and how often.
| Parameter | What it measures | Limit in IEC 61000-3-3 |
|---|---|---|
| Pst | Short-term flicker severity, how perceptible the flicker is over a short observation period | 1.0 |
| Plt | Long-term flicker severity, built up from a series of Pst values | 0.65 |
| dc | Relative steady-state voltage change, before and after a load change | 3.3 % |
| dmax | Maximum relative voltage change during an event, including switch-on | 4 %, or 6 % or 7 % for some manually switched, attended or infrequently switched equipment |
| d(t) / Tmax | How long the voltage change stays above a threshold during an event | No more than 500 ms above 3.3 % in a single voltage change |
What drives test duration
The standard sets fixed observation periods for Pst and Plt, and every operating mode that could produce flicker needs its own run. A product with a long operating cycle, several modes or a distinct switch-on behavior takes longer than one with a steady load. We plan the sequence from your operating modes when we quote.
How we perform harmonics and flicker testing
The product is powered from a stable, low-distortion AC source, so any distortion we measure is the product’s and not the supply’s. For flicker, a reference impedance is placed between the source and the product to represent a typical public supply network. A harmonic and flicker analyzer records the current and voltage throughout.
- Confirm the class and standard. We review your product and its ratings, and agree the applicable edition, the class and the operating modes.
- Set up on the reference source. The product connects to the AC source and, for flicker, the reference impedance, at its rated voltage on the 230 V class, 50 Hz supply the standards are written for.
- Run every relevant mode. We measure in the operating modes the standard calls for, including start-up and switching between modes where they matter.
- Evaluate against the limits. Harmonic currents are compared with the class limits, and Pst, Plt, dc, dmax and the time above threshold with the IEC 61000-3-3 limits.
- Report for the technical file. You receive an accredited test report ready for your technical file and Declaration of Conformity.
Harmonics and flicker standards we test to
Some of the standards we test to include:
| Standard | What it covers | Typical products |
|---|---|---|
| IEC 61000-3-2 | Harmonic current emissions, equipment up to 16 A per phase | Appliances, IT equipment, lighting, tools |
| EN 61000-3-2 | EU harmonized and UK designated version for CE and UKCA | Products sold in Europe and the UK |
| IEC 61000-3-3 | Voltage changes, fluctuations and flicker, up to 16 A per phase | Heaters, compressors, motor-driven equipment |
| EN 61000-3-3 | EU harmonized and UK designated version for CE and UKCA | Products sold in Europe and the UK |
| AS/NZS 61000-3-2 and 61000-3-3 | Australian and New Zealand adoptions | Where a customer or utility specifies them |
| SANS 61000-3-2 and 61000-3-3 | South African adoptions | Where a customer or utility specifies them |
These tests are usually called up by a product or generic standard such as IEC 61000-6-3, IEC 61326-1 or IEC 60601-1-2, which we also test to. Our testing is ISO/IEC 17025 accredited by A2LA under certificate 5179.02; see our accreditations.
Technical specifications
- AC source and analyzer system compliant with IEC 61000-4-7 for harmonics and IEC 61000-4-15 for flicker, with a flicker reference impedance
- Test supply calibrated at 230 V / 50 Hz
- Field testing of current harmonics and flicker at your site
On-site harmonics and flicker testing
Our accreditation includes field testing for current harmonics and for voltage fluctuations and flicker. Large machines, installed systems and equipment that cannot be shipped can be measured where they operate, and we agree the site arrangements with you when we scope the test.
Why products fail, and how to fix them
- No or weak power factor correction. A plain rectifier and bulk capacitor draw current in narrow peaks, which fails the harmonic limits, especially in Class C and D.
- Inrush at switch-on. A large charge current at power-up can push dmax over the limit.
- Heater and motor cycling. A thermostat or motor that switches a heavy load on and off at the wrong rate produces flicker.
- Firmware load steps. Control software that applies full power in one step, instead of ramping it, creates voltage changes the standard counts.
Most of these are fixed at design level, which is why an EMC pre-scan early in development costs far less than a failed formal test. When a product fails here, our engineers help identify the cause and retest.
Why test with Megalab
Accredited and recognized. ISO/IEC 17025 accredited by A2LA under certificate 5179.02, and FCC and ISED recognized for the rest of your EMC program.
On-site testing. Field testing for harmonics and flicker is part of our accreditation, which matters for machinery and installed equipment.
The rest of your CE EMC plan in the same lab. Conducted and radiated emissions, immunity testing including ESD and IEC 61000-4-11 voltage dips, and product safety all run at our lab in Aurora, Ontario.
What we need to quote you
- Product type, a short description and photos
- Rated input voltage, current and power; single-phase or three-phase
- Dimensions and weight
- Operating modes and the operating cycle, including how loads switch on and off
- Whether the product is lighting equipment
- Target markets, and the standard if you know it
- Samples available, and whether you want on-site testing or a pre-scan first
Send us the details and we will confirm the requirement and quote it.
Harmonics and flicker testing FAQ
Under the EMC Directive 2014/30/EU, equipment connected to the public low-voltage supply and rated up to 16 A per phase normally shows conformity by meeting the harmonized standards EN 61000-3-2 for harmonics and EN 61000-3-3 for flicker. The results go into the technical file that supports your Declaration of Conformity.
The product is powered from a stable AC source through a reference impedance that represents the public supply. A flicker analyzer records the voltage changes while the product runs through its operating modes, including switch-on. The results for Pst, Plt, dc, dmax and time above threshold are compared with the limits in the standard.
IEC 61000-3-2 limits the harmonic currents a product draws from the supply. IEC 61000-3-3 limits the voltage changes and flicker its load changes cause. Most mains products up to 16 A are assessed against both.
Lighting is Class C, portable tools are Class B, and personal computers, monitors, televisions and variable-speed refrigerators and freezers rated 600 W or less are Class D. Most other equipment is Class A. Send us the product details and we will confirm the class.
Pst is short-term flicker severity, how noticeable the flicker is to a person over a short period. Plt is long-term severity, built up from a series of Pst values.
Yes. Field testing for both is part of our accreditation, so equipment that cannot be shipped can be measured where it is installed.
Usually because the input stage has no power factor correction, so current is drawn in narrow peaks. Our engineers help identify the cause, and a pre-scan catches it early.
Ready to test? Send us your product details and we will confirm the harmonics and flicker requirement and quote it. Request a Harmonics and Flicker Quote.